Surface Roughness Tester Gauge Profile Tester 2 Parameters Ra Rz ProfilometerFeature:
* Be compatible with four standards of ISO, DIN, ANSI and JIS. When measuring the roughness of a surface, the sensor is placed on the surface and then uniformly slides along the surface by driving the mechanism inside the tester. The sensor gets the surface roughness by the sharp built-in probe. This roughness causes displacement of the probe which results in change of inductive amount of induction coils so as to generate analogue signal, which is in proportion to the surface roughness at output end of phase-sensitive rectifier. The exclusive DSP processes and calculates and then outputs the measurement results on LCD.
* Multiple parameter measurement: Ra, Rz.
* Highly sophisticated inductance sensor.
* Built-in lithium ion rechargeable battery and control circuit with high capacity.
* Small in size, light in weight and easy to use.
* Manual or automatic shut down.
* Metric /Imperial Conversion.
* Use “USB data output” and “RS-232 data output” to connect with PC.
(Optional Accessories)
* Provide “Bluetooth ™ data output” choice.
(Optional Accessories)
Specification:
Display
4 digits, 10 mm LCD, with blue backlight
Parameters
Ra
、
Rz
Measuring Range
Ra: 0.05~10.00 um / 1.000~400.0 uinch
Rz: 0.020~100.0 um / 0.780~4000 uinch
Accuracy
≤ ± 10%
Fluctuation of display value
Not more than 6%
Sensor
Test Principle
Inductance type
Radius of Probe Pin
10 um
Material of Probe Pin
Diamond
Measurement Force of Probe
16 mN
(
1.6 gf
)
Probe Angle
90°
Vertical Radius of Guiding Head
48 mm
Maximum driving stroke
17.5 mm / 0.7 inch
Cutoff length (l)
0.25 mm / 0.8 mm / 2.5 mm optional
Driving speed
Sampling
When Length = 0.25 mm, Vt = 0.135 mm/s
When Length = 0.8 mm, Vt = 0.5 mm/s
When Length = 2.5 mm, Vt = 1 mm/s
Returning
Vt = 1 mm/s
Resolution
0.001 um, if reading
<
10um
0.01 um, if 10um ≤ reading ≤ 100um
0.1 um, if reading ≥ 100um
Evaluation length
1~5 L optional
Auto Power Off
Operating conditions
Temperature: 0~50
℃
Humidity:
<
80 %RH
Power Supply
Built-in rechargeable Li-ion battery
Dimensions
140 × 52 × 48 mm
Weight
420 g
Standard Accessories :
Main Unit, Standard Sample Plate SSP-100, Standard Sensor SSP-100 Small Screwdriver, Power Adapter, Carrying Case ( B09 ), Operation Manuall